Power
SCPI Commands :
CONFigure:GPRF:MEASurement<Instance>:POWer:MODE
CONFigure:GPRF:MEASurement<Instance>:POWer:TOUT
CONFigure:GPRF:MEASurement<Instance>:POWer:SLENgth
CONFigure:GPRF:MEASurement<Instance>:POWer:MLENgth
CONFigure:GPRF:MEASurement<Instance>:POWer:REPetition
CONFigure:GPRF:MEASurement<Instance>:POWer:SCOunt
CONFigure:GPRF:MEASurement<Instance>:POWer:PDEFset
- class PowerCls[source]
Power commands group definition. 54 total commands, 5 Subgroups, 7 group commands
- get_mlength() float [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:MLENgth value: float = driver.configure.gprf.measurement.power.get_mlength()
Sets the length of the evaluation interval used to measure a single set of current power results. The measurement length cannot be greater than the step length.
- return:
meas_length: No help available
- get_mode() CcdfMode [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:MODE value: enums.CcdfMode = driver.configure.gprf.measurement.power.get_mode()
Selects the measurement mode for measurements without list mode. Select the mode before starting the power measurement.
- return:
ccdf_mode: No help available
- get_pdef_set() str [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:PDEFset value: str = driver.configure.gprf.measurement.power.get_pdef_set()
No command help available
- return:
predefined_set: No help available
- get_repetition() Repeat [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:REPetition value: enums.Repeat = driver.configure.gprf.measurement.power.get_repetition()
Specifies the repetition mode of the measurement. The repetition mode specifies whether the measurement is stopped after a single shot or repeated continuously. Use CONFigure:..:MEAS<i>:…:SCOunt to determine the number of measurement intervals per single shot.
- return:
repetition: SINGleshot: single-shot measurement CONTinuous: continuous measurement
- get_scount() int [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:SCOunt value: int = driver.configure.gprf.measurement.power.get_scount()
Specifies the statistic count of the measurement. The statistic count is equal to the number of measurement intervals per single shot.
- return:
statistic_count: Number of measurement intervals
- get_slength() float [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:SLENgth value: float = driver.configure.gprf.measurement.power.get_slength()
Sets the time between the beginning of two consecutive measurement lengths.
- return:
step_length: No help available
- get_timeout() float [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:TOUT value: float = driver.configure.gprf.measurement.power.get_timeout()
Defines a timeout for the measurement. The timer is started when the measurement is initiated via a READ or INIT command. It is not started if the measurement is initiated manually. When the measurement has completed the first measurement cycle (first single shot) , the statistical depth is reached and the timer is reset. If the first measurement cycle has not been completed when the timer expires, the measurement is stopped. The measurement state changes to RDY. The reliability indicator is set to 1, indicating that a measurement timeout occurred. Still running READ, FETCh or CALCulate commands are completed, returning the available results. At least for some results, there are no values at all or the statistical depth has not been reached. A timeout of 0 s corresponds to an infinite measurement timeout.
- return:
tcd_timeout: No help available
- set_mlength(meas_length: float) None [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:MLENgth driver.configure.gprf.measurement.power.set_mlength(meas_length = 1.0)
Sets the length of the evaluation interval used to measure a single set of current power results. The measurement length cannot be greater than the step length.
- param meas_length:
No help available
- set_mode(ccdf_mode: CcdfMode) None [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:MODE driver.configure.gprf.measurement.power.set_mode(ccdf_mode = enums.CcdfMode.POWer)
Selects the measurement mode for measurements without list mode. Select the mode before starting the power measurement.
- param ccdf_mode:
POWer: Power mode STATistic: Statistic mode
- set_pdef_set(predefined_set: str) None [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:PDEFset driver.configure.gprf.measurement.power.set_pdef_set(predefined_set = 'abc')
No command help available
- param predefined_set:
No help available
- set_repetition(repetition: Repeat) None [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:REPetition driver.configure.gprf.measurement.power.set_repetition(repetition = enums.Repeat.CONTinuous)
Specifies the repetition mode of the measurement. The repetition mode specifies whether the measurement is stopped after a single shot or repeated continuously. Use CONFigure:..:MEAS<i>:…:SCOunt to determine the number of measurement intervals per single shot.
- param repetition:
SINGleshot: single-shot measurement CONTinuous: continuous measurement
- set_scount(statistic_count: int) None [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:SCOunt driver.configure.gprf.measurement.power.set_scount(statistic_count = 1)
Specifies the statistic count of the measurement. The statistic count is equal to the number of measurement intervals per single shot.
- param statistic_count:
Number of measurement intervals
- set_slength(step_length: float) None [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:SLENgth driver.configure.gprf.measurement.power.set_slength(step_length = 1.0)
Sets the time between the beginning of two consecutive measurement lengths.
- param step_length:
No help available
- set_timeout(tcd_timeout: float) None [source]
# SCPI: CONFigure:GPRF:MEASurement<Instance>:POWer:TOUT driver.configure.gprf.measurement.power.set_timeout(tcd_timeout = 1.0)
Defines a timeout for the measurement. The timer is started when the measurement is initiated via a READ or INIT command. It is not started if the measurement is initiated manually. When the measurement has completed the first measurement cycle (first single shot) , the statistical depth is reached and the timer is reset. If the first measurement cycle has not been completed when the timer expires, the measurement is stopped. The measurement state changes to RDY. The reliability indicator is set to 1, indicating that a measurement timeout occurred. Still running READ, FETCh or CALCulate commands are completed, returning the available results. At least for some results, there are no values at all or the statistical depth has not been reached. A timeout of 0 s corresponds to an infinite measurement timeout.
- param tcd_timeout:
No help available
Cloning the Group
# Create a clone of the original group, that exists independently
group2 = driver.configure.gprf.measurement.power.clone()
Subgroups